Image Optimization
5-5
Display
F 5.0
D 17.6
G 50
FR 40
IP 4
DR 60
Parameter Frequency Depth Gain
Frame Rate
B IP
B Dynamic Range
Parameters that can be adjusted to optimize the B Mode image are indicated in the
following.
Adjustment
Items
Control Panel
Gain, Depth, TGC, iTouch
Menu and Soft
Menu
Dynamic Range, Focus Number, FOV Position, Line Density, IP,
Colorize, L/R Flip, Rotation, Persistence, Colorize Map, U/D Flip,
iTouch, Frequency, Gray Map, Focus Position, iClear, FOV,
Smooth, TSI, Curve, Gray Rejection,
γ
, High FR, iTouch Bright, A.
power, B Steer, iBeam, Trapezoid, Image Merge, ExFOV, HS Scale
Items that appear in the menu or the soft menus are dependent upon preset, which
can be changed or set through "[Setup]
→
[Image Preset]"; please refer to "5.13
Image Preset" for details.
5.3.3 B Mode Image Optimization
Gain
Description
To adjust the gain of the whole receiving information in B mode. The
real-time gain value is displayed in the image parameter area in the upper
left corner of the screen.
Operation
Rotate the <B> knob clockwise to increase the gain, and anticlockwise to
decrease.
Or adjust it in the image parameter area.
The adjusting range is 0-100.
Effects
Increasing the gain will brighten the image and you can see more received
signals. However, noise may also be increased.
Depth
Description
This function is used to adjust the display depth of sampling, the real-time
value of which is displayed on the image parameter area in the upper left
corner of the screen.
Operation
Rotate the “Depth” knob clockwise to increase the depth;
Rotate the “Depth” knob anticlockwise to decrease the depth.
The adjustable depth values vary depending upon the probe types.
Effects
Increase the depth to see tissue in deeper locations, while decrease the
depth to see tissue in shallower locations.
Impacts
Depth increase will cause a decrease in the frame rate.
TGC
Description
The system compensates the signals from deeper tissue by segments to
optimize the image.
There are 8-segment TGC sliders on the control panel corresponding to the
areas in the image.
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...