DICOM
12-3
12.1.2 DICOM Local Setting
To set the DICOM server properties and log saving manner.
1. Press <Setup> to enter the Setup menu.
2. Move the cursor onto [Network Preset], select [DICOM Local] to open the screen, as
shown in the figure below:
3. DICOM Local Setting items are described as follows:
Name
Description
AE Title
Application entity title of the ultrasound system.
The AE title here should be the same with the one of the acceptable SCU set
in the server.
Port
DICOM communication port, which should be the same with the one in the
server.
PDU
Maximum PDU data package size ranging from 16384 to 65536. The default
value is 32768.
Log
Setting
File Name: name of the DICOM log, generated by the system automatically.
Log Level: to set the log type.
The available selections: No log, Service result log, Service process log, All
log.
We suggest setting “All log”, so that you can see the DICOM log on the E
disk.
OK
Click to confirm parameter setting.
Cancel
Click to cancel parameter setting.
Enter AE Title, Port, PDU according to the actual situation, and then click [OK] to exit the
screen.
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...