System Maintenance 19-5
No.
Failure
Cause
Measure
3
The monitor
displays the
characters and
menus but no
images.
The
transmission
power, overall
gain or TGC
controls are
improperly set.
Check if a probe
is connected
and/or fully
connected.
The system is in
the frozen status.
Adjust the transmission power,
gain or TGC control.
Ensure proper probe
connection.
Unfreeze the image.
4
The image
quality is
degraded
The exam mode
is not correct.
The
settings
of
the image post
processing are
not correct.
The
image
presets are
inappropriate.
Select an appropriate exam
mode.
Adjust the settings of the image
via post processing or reset the
post processing to the default
values.
Reset the factory default
presets.
5
The button is
responseless
with the system
buzzing
There is too
much dirt
blocking the
button
Check the control panel for the
button being blocked and press
it several times to release it.
Clean the button.
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...