5-14 Image Optimization
Click [
γ
] in the soft menu or menu to adjust.
The adjusting range is 0-3.
Impacts
The function is available in real-time imaging, freeze or cine review status.
The post process adjustment will not influence the cine review.
Gray Map
Description
This function applies the gray correction to obtain the optimum image
through adjusting 3 parameters: curve, gray rejection and
γ
.
Operation
Select among the maps through the [Gray Map] item in the soft menu or
menu.
You can also adjust through the grayscale bar: move the cursor to the
grayscale bar and press <Set> on the control panel to adjust.
The system provides 8 groups of gray map.
Impacts
The function is available in real-time imaging, freeze or cine review status.
Edge Enhance
Description
This function is used to increase image profile, so as to distinguish the
image boundary.
Operation
Adjust through the [Edge Enhance] item in the soft menu or menu.
The system provides 4 levels of edge enhance effects, off represents no
edge enhance is turned on, and the bigger the value the stronger the effect.
Impacts
Larger edge enhance may lead to noise increasing.
Dynamic Range
Description
This function is used to adjust the B image resolution to compress or
expand the gray display range. The real-time dynamic range value will be
displayed in the image parameter area in the upper left corner of the
screen.
Operation
Adjust through the [Dyn Ra.] item in the soft menu or menu.
Or adjust it in the image parameter area.
The adjusting range is 30-160dB in increments of 5.
Effects
The more the dynamic range, the more specific the information, and the
lower the contrast with more noise.
M Soften
Description
This feature is used to process the scan lines of M images to reject noise,
making the image details to be clearer.
Operation
Adjust through the [M Soften] item in the soft menu or menu.
The system provides 5 levels of M Soften adjustment, the bigger the value
the stronger the effect.
5.5
Color Mode Image Optimization
The Color mode is used to detect color flow information, and the color is designed to judge
the direction and speed of blood flow.
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...