7-8 Display & Cine Review
z
There is sound feedback after a saving is completed; meanwhile, a thumbnail is
showed in the Thumbnail area. If the saving fails, please pay attention to the help
information.
z
If you perform any operation that will clear the cine memory during the cine
saving process, the cine saving is cancelled. (Refer to 7.6 Cine Memory for
details).
7.6 Cine
Memory
7.6.1 Cine Memory Setting
There are 2 ways of cine memory split: auto and split.
Setting path: select "Auto" or "Split" for cine memory in [Setup]
→
[System Preset]
→
[Image Preset].
Where,
z
"Auto" for the cine memory indicates the system splits the cine memory as per the
number of B image windows.
z
“Split” indicates the system always splits the cine memory, even if in the single-B
window, the system splits the cine memory into two as well. You can press <B>
key to switch and display the images in either memory, so you can compare
images in Split mode.
The memory capacity is evenly distributed as per splitting number, as shown in the
following table (Taking B images in low density as an example, the capacity of B cine
memory is N frames):
Imaging Mode
Split
Single-B/ Color
Dual
Quad
Auto
One memory, with
capacity N frames.
The memory splits into
two, with capacity N/2
frames each
The memory splits into
four, with capacity N/4
frames each
Split
The memory splits
into two, with
capacity N/2 frames
each
The memory splits into
two, with capacity N/2
frames each
The memory splits into
four, with capacity N/4
frames each
7.6.2 Cine Memory Clear
In the following conditions, the cine review memory will be cleared:
z
Start an exam of a new patient.
z
Start a new exam for the same patient.
z
Switching the probe (if the cine memory is split, only the cine memory
corresponding to the currently activated window is cleared)
z
Changing the exam condition (if the cine memory is split, only the cine memory
corresponding to the currently activated window is cleared)
z
Switching in some imaging mode: include imaging mode switching among B, M,
Color, PW, Power, rescanning imaging in iScape, and image re-acquisition in
3D/4D, display mode switching etc.
z
Parameters modification, including:
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...