13.2.6 iNeedle
In the course of biopsy, the metal needle attached to the probe is punctured into the tissue
with a certain angle; because of the needle acoustic impedance, the ultrasonic beam
cannot penetrate the metal needle, a reflecting boundary is formed. As in figure 1, if the
deflection angle is very large, the needle display is not clear.
In the condition of deflected ultrasonic transmission, the beam direction is perpendicular to
the needle direction, and the reflection direction will be the same with the needle, as
shown in figure 2, when the needle display in the ultrasound image is very clear. The
system provides an additional deflection transmission that is approximately perpendicular
to the metal needle, as the normal transmission (perpendicular to the transducer surface)
is contained as well. And the deflection angel can be chosen by users.
Figure 1 Figure 2
To enter/exit iNeedle
To enter iNeedle
Click [iNeedle] item in B soft menu.
Or, you can assign a user-defined key for entering iNeedle.
Open iNeedle in Biopsy status
1. Perform scanning and locate the target, press <Biopsy> to enter the screen.
2. Click [iNeedle] to enter the status, available adjusting parameters are displayed on
the menu.
To exit iNeedle
Press the user-defined key or click [iNeedle] to exit the status and enter B mode.
Needle
Steer
Description
This function adjusts the biopsy needle angle via changing the steer angle
of the scan line. The iNeedle affecting region changes correspondingly.
Operation
Adjust through the [Needle Steer] item on the menu.
B/iNeedle
Description
This function is used to display B image and iNeedle image synchronously.
Operation
Adjust through [B/iNeedle] on the menu.
Tips: iZoom (full-screen magnification) is available on iNeedle status.
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...