8
Setting Measurement Conditions
Update rate of each module
The data refresh rate is not allowed to exceed the maximum sampling rate of each module.
The instrument measures the same data until the data gets updated, causing the instrument to plot
a stair-step waveform.
In addition, even though the instrument measures the same signal simultaneously, values may vary
due to differences in the sampling rate, frequency range, and frequency characteristics of modules.
Modules
Maximum sampling rate
or data update rate
Reference
Model 8966 Analog Unit
20 MS/s (50 ns)
–
Model 8967 Temp Unit
Depends on the data refresh setting.
Model 8968 High Resolution Unit
1 MS/s (1 µs)
–
Model U8969 Strain Unit
200 kS/s (5 µs)
–
Model 8970 Freq Unit
Depends on the setting.
Model 8971 Current Unit
1 MS/s (1 µs)
–
Model 8972 DC/RMS Unit
Depends on the response setting.
Model 8973 Logic Unit
20 MS/s (50 ns)
–
MR8990
Digital Voltmeter Unit
Depends on the NPLC setting.
Model U8974 High Voltage Unit
Depends on the response setting.
Model U8975 4ch Analog Unit
5 MS/s (200 ns)
–
Model U8976 High Speed Analog Unit
200 MS/s (5 ns)
–
Model U8977 3CH Current Unit
5 MS/s (200 ns)
–
Model U8978 4CH Analog Unit
5 MS/s (200 ns)
–
U8979 Charge Unit
200 kS/s (5 µs)
–
HIOKI MR6000A966-03
Summary of Contents for Memory HiCorder MR6000
Page 2: ...HIOKI MR6000A966 03...
Page 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Page 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Page 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Page 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Page 391: ...HIOKI MR6000A966 03...
Page 392: ...HIOKI MR6000A966 03...