128
Setting the Trigger Satisfaction Conditions (AND/OR Operation) Among Trigger Sources
Setting example: To trigger the instrument when a waveform crosses the zero-volt level
while rising (
[ ]
*)
The instrument is triggered based on whether the trigger satisfaction condition is set to AND or OR
operation in the following ways:
Channel
Trigger
Level
Slope
Filter
CH1, CH2
Level
0.00 V
*
Off
[AND]
[OR]
One waveform is above 0 V, and
the other is also above 0 V.
Either waveform crosses 0 V
upwards.
Start
Start
When the trigger timing is set to
[Start/Stop]
, the instrument determines that a logical AND or OR
operation is satisfied in a set of trigger sources assigned to the start trigger or those assigned to the
stop trigger.
*: With the
[AND]
setting, the slope setting item appears as
[HIGH]
.
HIOKI MR6000A966-03