172
Configuring the Numerical Calculation Settings
Setting
Description
Sample screen
Burst width
Target channel
(Analog, logic, real-time
waveform calculation)
Allows you to specify
calculation target channels.
Slope
(Logic channels
only)
Detects rising edges and
calculates a burst width.
Detects falling edges and
calculates a burst width.
Filter(samples)
(
Off
,
10
to
10,000
)
Allows you to enter duration
used to determine that a
waveform crosses the specified
level.
Only after a waveform has
crossed the level and does not
cross the level again within
the specified filter duration,
the waveform is considered to
have crossed the level.
This is useful to eliminate false
level-crossing events caused
by noise.
Burst end filter(samples)
(
Off
,
10
to
10,000
)
Allows you to set duration
used to determine whether a
waveform is a burst signal.
If a period from the time when
a waveform falls within the
window level range to the
time when it exits the range
becomes longer than the
specified filter duration, the
instrument determines the
detected period to be a burst
duration.
Window-level range*
(
Up
,
Low
)
Allows you to enter an upper
and lower limit values used to
determine whether a waveform
is a burst signal.
Stat.
Top
Detects the value obtained first
from the beginning within the
specified range.
Average
Calculates the average,
maximum, or minimum value
for each parameter within the
specified range.
Max
Min
Times
Calculates the value when the
number of times a waveform
crosses the level reaches the
specified number from the
beginning within the specified
range.
*: You can set a level within the range of −9.9999E+29 to −1.0000E29, at zero, within the range of +1.0000E29 to
+9.9999E+29. A value to five significant figures can be specified (a value to 10 significant figures can be specified
for a time of the level at time setting).
HIOKI MR6000A966-03
Summary of Contents for Memory HiCorder MR6000
Page 2: ...HIOKI MR6000A966 03...
Page 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Page 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Page 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Page 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Page 391: ...HIOKI MR6000A966 03...
Page 392: ...HIOKI MR6000A966 03...