135
Triggering the Instrument Using Analog Signals (Analog Trigger)
4.
[Period-in]
trigger and
[Period-out]
trigger
The instrument measures periods of an input waveform, which are time lags between consecutive
two points at which an input voltage crosses the specified level in the positive or negative direction.
The period-trigger condition is satisfied when a period is inside the specified range (In) or outside
the specified range (Out).
An trigger point lags behind the actual trigger point by one sample. These triggers are disabled
when the sampling rate is set at 200 MS/s.
The external sampling is cannot be set when the envelope is used. These triggers cannot be set
either when Model MR8990 and Model 8970 is used.
See “Setting of the period range” and “[Period-out] trigger” (p. 136).
Level
Period upper limit
Period lower limit
Upper
limit value
Out of the range
Within the period range
Setting
Description
Level
–f.s.
to
+f.s.
Default:
0
Allows you to enter a level for detecting the
rising or falling slopes of a signal.
Slope
,
Allows you to choose which of the following
points to use to calculate periods: two
consecutive points at which a signal crosses
the specified level while rising; or those while
falling.
Period lower
limit
*
Zero or 5 times the sampling period
or more
The lower limit of periods cannot be set to a
value higher than that in the
[Period upper
limit]
box. When the lower limit of periods is
set to
[0]
, the instrument ignores the value in
the
[Period lower limit]
box, and the period-
trigger condition is satisfied using the value in
the
[Period upper limit]
box only.
Period upper
limit
*
20,000 times or less of the sampling
period
The higher limit of periods cannot be set to
a value lower than that in the
[Period lower
limit]
box.
Event
With
OR
1
to
4,000
Allows you to enter the number of events.
The instrument counts the number of times
the period-trigger condition is satisfied. Only
after the count reaches the specified number of
events, an analog trigger is generated.
With
AND
Not available
Filter
Off
,
10
to
10,000
Allows you to enter a filter width in the number
of samples. Only after the period-trigger
condition is continuously satisfied during the
specified period, an analog trigger is generated.
Specifying this option prevents the instrument
from being unintentionally triggered due to
noise.
*: Configurable ranges of the period lower limit and period upper limit vary according to the sampling rates
(periods).
Configuring the
Trigger Settings
HIOKI MR6000A966-03
Summary of Contents for Memory HiCorder MR6000
Page 2: ...HIOKI MR6000A966 03...
Page 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Page 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Page 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Page 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Page 391: ...HIOKI MR6000A966 03...
Page 392: ...HIOKI MR6000A966 03...