72
Configuring Module-Specific Settings
10
Tap the
[Smoothing]
box, and then choose a scaling setting from the list.
Only when
[Mode]
is set to
[Freq]
or
[Revolution]
, this setting is available.
Off
Records measured data without smoothing (resulting in a step-like waveform).
On
Interpolates measured data to smooth a waveform and outputs it.
(Upper limit: 10 kHz; outputting data with this setting set to on lags behind that with this
setting set to off)
11
Tap the
[Hold]
box, and then choose a measured-value retaining setting from the list.
Only when
[Mode]
is set to
[Freq]
or
[Revolution]
, this setting is available.
Off (1 Hz)
,
Off (0.5 Hz)
,
Off (0.2 Hz)
,
Off (0.1 Hz)
When the instrument cannot determine a measured value even when the frequency
reaches a value in the brackets, the instrument determines the measurement to stop
and regards the measured value as 0 Hz (0 rpm).
On
Retains the value confirmed the last time.
12
Tap
[Close]
.
The setting dialog box closes.
HIOKI MR6000A966-03