83
Configuring Module-Specific Settings
When using an out-of-setting-range current sensor
You can use an out-of-setting-range current sensor using the scaling function.
See “Automatically saving waveform data” (p. 102).
1
Tap the area that includes
[Sensitivity]
.
The setting dialog box appears.
2
Tap
[Sensitivity]
box, and then enter sensor sensitivity.
Multiply the sensor sensitivity of a sensor to be used by a certain value to allow a product to fall within the
setting range (0.1 to 10), and enter the product.
3
Tap
[Close]
.
The setting dialog box closes.
4
Tap the
[Scaling]
box, and then configure the scaling setting.
Configure the scaling setting so that a scaling ratio is the same value as the number you multiplied the sensor
sensitivity by.
Example 1
For sensor sensitivity of 23.4 pC/(m/s
2
)
Specify 10 pC/(m/s
2
), which results from multiplying the sensor sensitivity by 1/2.34, as the sensor sensitivity.
To display measured values after multiplying them by 1/2.34, configure the scaling setting as follows:
To configure the scaling setting using the conversion ratio method
Ratio
0.4274E+00 (= 10/23.4)
Offset
0.0000E+00
Units
m/s
2
To configure the scaling setting using the 2-point method
Input1
2.3400E+00
Scale1
1.0000E+00
Input2
0.0000E+00
Scale2
0.0000E+00
Units
m/s
2
Example 2
For sensor sensitivity of 0.05 pC/(m/s
2
)
Specify 0.1 pC/(m/s
2
), which results from multiplying the sensor sensitivity by two, as the sensor sensitivity.
To display measured values after multiplying them by two, configure the scaling setting as follows:
To configure the scaling setting using the conversion ratio method
Ratio
2.0000E+00 (= 0.1/0.05)
Offset
0.0000E+00
Units
m/s
2
To configure the scaling setting using the 2-point method
Input1
0.0500E+00
Scale1
1.0000E+00
Input2
0.0000E+00
Scale2
0.0000E+00
Units
m/s
2
Advanced Functions
HIOKI MR6000A966-03
Summary of Contents for Memory HiCorder MR6000
Page 2: ...HIOKI MR6000A966 03...
Page 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Page 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Page 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Page 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Page 391: ...HIOKI MR6000A966 03...
Page 392: ...HIOKI MR6000A966 03...