220
Examples of Configuring Waveform Calculation Settings
Table A
Relationship among sampling rates, sampling frequencies, cutoff
frequencies, and setting resolutions of cutoff frequencies (for both
FIR and IIR filters)
Sampling rate
Sampling
period
Sampling
frequency
Upper limit of cutoff
frequency
(Value less than this
frequency can be set)
Cutoff frequency
setting resolution
200 MS/s
5 ns
200 MHz
100 MHz
200 kHz
100 MS/s
10 ns
100 MHz
50 MHz
100 kHz
50 MS/s
20 ns
50 MHz
25 MHz
50 kHz
20 MS/s
50 ns
20 MHz
10 MHz
20 kHz
10 MS/s
100 ns
10 MHz
5 MHz
10 kHz
5 MS/s
200 ns
5 MHz
2.5 MHz
5 kHz
2 MS/s
500 ns
2 MHz
1 MHz
2 kHz
1 MS/s
1 µs
1 MHz
500 kHz
1 kHz
500 kS/s
2 µs
500 kHz
250 kHz
500 Hz
200 kS/s
5 µs
200 kHz
100 kHz
200 Hz
100 kS/s
10 µs
100 kHz
50 kHz
100 Hz
50 kS/s
20 µs
50 kHz
25 kHz
50 Hz
20 kS/s
50 µs
20 kHz
10 kHz
20 Hz
10 kS/s
100 µs
10 kHz
5 kHz
10 Hz
5 kS/s
200 µs
5 kHz
2.5 kHz
5 Hz
2 kS/s
500 µs
2 kHz
1 kHz
2 Hz
1 kS/s
1 ms
1 kHz
500 Hz
1 Hz
500 S/s
2 ms
500 Hz
250 Hz
500 mHz
200 S/s
5 ms
200 Hz
100 Hz
200 mHz
100 S/s
10 ms
100 Hz
50 Hz
100 mHz
50 S/s
20 ms
50 Hz
25 Hz
50 mHz
20 S/s
50 ms
20 Hz
10 Hz
20 mHz
10 S/s
100 ms
10 Hz
5 Hz
10 mHz
5 S/s
200 ms
5 Hz
2.5 Hz
5 mHz
2 S/s
500 ms
2 Hz
1 Hz
2 mHz
1 S/s
1 s
1 Hz
0.5 Hz
1 mHz
• Stating a measurement and waveform calculation with the instrument having no waveform data
allows the sampling rate setting on the measurement setting screen or the waveform screen to
define a settable upper limit of a cutoff frequency.
• Tapping
[Execute]
on the waveform calculation screen to perform calculations using waveform
data previously acquired in the instrument allows the sampling rate setting to define a settable
upper limit of a cutoff frequency.
• The instrument with the external sampling clock setting performs calculations with one sample
fixed at 1 Hz.
HIOKI MR6000A966-03
Summary of Contents for Memory HiCorder MR6000
Page 2: ...HIOKI MR6000A966 03...
Page 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Page 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Page 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Page 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Page 391: ...HIOKI MR6000A966 03...
Page 392: ...HIOKI MR6000A966 03...