Evaluating Calculation Results on a Pass/Fail Basis
> [Calculation] > [Numerical calculation]
4
3
1
2
1
Tap the
[Judge stop condition]
box, and then from the list, choose a stopping condition
applied according to judgments.
PASS
Stops measurement when a calculation result falls within a criteria range (pass
judgment).
FAIL
Stops measurement when a calculation result is outside a criteria range (fail judgment).
PASS & FAIL
Stops measurement regardless whether a pass or fail judgment is given.
2
Tap the
[Judge]
box, and then from the list, choose whether or not to evaluate calculation
results.
Off
Evaluates calculation results.
On
Yields a FAIL judgment if a result is outside a criteria range.
Calculated values in a channel with the FAIL judgment are highlighted in red.
3
Tap the
[Up]
and
[Low]
boxes in turn, and then enter evaluation criteria values in each of
them.
Enterable range: −9.9999E+29 to -1.0000E29, 0, +1.0000E29 to +9.9999E+29
4
(To evaluate existing data)
Tap
[Execute]
.
(To automatically execute evaluation after measurement)
START
key
HIOKI MR6000A966-03