171
Configuring the Numerical Calculation Settings
Setting
Description
Sample screen
+Width
−Width
Target channel
(Analog, real-time
waveform calculation)
Allows you to specify
calculation target channels.
Filter(samples)
(
Off
,
10
to
10,000
)
Allows you to enter duration
used to determine that a
waveform crosses the specified
level.
Only after a waveform has
crossed the level and does not
cross the level again within
the specified filter duration,
the waveform is considered to
have crossed the level.
This is useful to eliminate false
level-crossing events caused
by noise.
Stat.
Top
Detects the value obtained first
from the beginning within the
specified range.
Average
Calculates the average,
maximum, or minimum value
for each parameter within the
specified range.
Max
Min
Times
Calculates the value when the
number of times a waveform
crosses the level reaches the
specified number from the
beginning within the specified
range.
Numerical Calculation Function