141
Triggering the Instrument at Regular Intervals (Interval Trigger)
Acquiring data at regular intervals (relation between a time interval and a
recording length or recording time)
The instrument is not triggered until having acquired data that has the specified recording length or
recording interval.
When a recording length or recording time is
shorter than a time interval
When a recording length or recording time is
longer than a time interval
The instrument records data
that has the specified recording
length or recording time.
Start
Time interval
Finish
Start
Time interval
Finish
The instrument records data
that has the specified recording
length or recording time.
Configuring the
Trigger Settings
HIOKI MR6000A966-03