62
Fine-Adjusting Input Values (Vernier Function)
3.4 Fine-Adjusting Input Values (Vernier Function)
You can freely fine-adjust input voltage on the waveform screen. When recording physical values,
such as noise, temperature, and acceleration, with sensors, you can adjust those amplitudes, which
facilitates calibration.
Normal display
Waveform processed
by the vernier function
When an input voltage of 1.2 V is
displayed as a voltage of 1.0 V
1.0 V
1.2 V
> [Channel]
1
Tap
[Vernier]
.
The adjustment keypad appears.
2
While observing the waveform, you can fine-adjust its amplitude by tapping
[−−]
,
[−]
,
[+]
, and
[++]
.
Tap
[C]
to revert the fine -adjusted amplitude to its original.
•
The adjustable range is from 50% to 250% of an original waveform.
• Using the vernier function turns the sides of
[Vernier]
blue.
Waveform not processed by the vernier function Waveform processed by the vernier function
•
Waveform data (data saved as files) is that adjusted by the vernier function.