261
11
Dual Sampling
Dual sampling is a function that performs high-speed sampling in combination with low-speed
sampling; the high-speed sampling can capture abnormal phenomena using triggers to record
instantaneous waveforms while the low-speed sampling is continuously recording normal-state
trend waveforms.
Multiple abnormal phenomena can be recorded using the memory division function.
Trend waveform
By observing trend waveforms, you can roughly check variation of waveforms under measurement.
Instantaneous waveforms can be captured by configuring trigger settings, allowing you to examine
abnormal phenomena (circled) closely.
Instantaneous waveform
Block 1
Block 2
Block 3
Block 4
Block 5
By pressing the
START
key, you can start recording trend waveforms. When trigger conditions are
met, instantaneous waveform recording can starts.
After measurement has finished, you can use instantaneous waveforms for numerical calculation,
waveform calculation, and FFT calculation.
11 Dual Sampling
Dual Sampling
HIOKI MR6000A966-03