300
Configuring the Transmission Function Settings (Creating a Transmission List)
Settings of the transmission list
No.
Name
Function
4
Use
Allows you to choose whether to transmit the data concerned.
Off
Does not transmit the data.
On
Transmit the data.
5
Timing
Allows you to set when to transmit the data concerned.
Key S1
When you press the
S1
key
Key S2
When you press the
S2
key
Start
When you press the
START
key
Otherwise, when you send the
:START
command
to start measurement
Trigger
When the trigger activates
Reply
When the instrument receives a message ID
set in
[Reply ID]
during measurement. (The
instrument does not reply any response while no
measurement is being performed.)
Pass
When a pass judgment is yielded for the
numerical calculation
Fail
When a fail judgment is yielded for the numerical
calculation
Error
When an error occurs or a warning is issued
6
Type
Allows you to choose a data type to be transmitted.
CAN Std.
CAN data frame in the base format
CAN Ext.
CAN data frame in the extended format
CAN Remote Std.
CAN remote frame in the base format
CAN Remote Ext.
CAN remote frame in the extended format
CAN FD Std.
CAN FD data frame in the base format
CAN FD Ext.
CAN FD data frame in the extended format
Some types of data cannot be sent depending on the CAN mode-setting
•In CAN mode, data set to
[CAN FD Std.]
or
[CAN FD Ext.]
cannot be
sent.
•In CAN FD mode, data set to
[CAN Remote Std.]
or
[CAN Remote Ext.]
cannot be sent.
7
Ch
Allows you to specify a CAN channel for use in transmission.
C1
,
C2
,
C3
,
C4
,
All
When you choose
[All]
, the instrument transmits the same data for all
channels.
8
Send ID
Allows you to enter a message ID of the data to be transmitted. Use
hexadecimal notation.
For the base format
0
to
7FF
For the extended format
0
to
1FFFFFFF
HIOKI MR6000A966-03
Summary of Contents for Memory HiCorder MR6000
Page 2: ...HIOKI MR6000A966 03...
Page 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Page 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Page 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Page 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Page 391: ...HIOKI MR6000A966 03...
Page 392: ...HIOKI MR6000A966 03...