209
Examples of Configuring Waveform Calculation Settings
8.5
Examples of Configuring Waveform Calculation
Settings
Obtaining an RMS waveform from an instantaneous waveform
This section explains how to display a RMS waveform obtained through a calculation using a
constantly oscillating waveform acquired across the analog channel (CH1-1). The instrument
provides no calculations to obtain an RMS waveform from a waveform with a variable period.
This example demonstrates a calculation for a waveform data with a period of 200 samples.
> [Calculation] > [Waveform calculation]
2
3
1
7
4
1
Tap the
[Calculation]
button to set it to
[On]
.
2
Tap the
[Area]
box, and then choose
[Whole]
.
3
Tap the channel to use for calculation,
[Z1]
, to set it to
[On]
.
4
Specify an arithmetic expression.
Tap the
[Formula]
box of
[Z1]
to open the arithmetic expression setting dialog box.
W
aveform Calculation Function
HIOKI MR6000A966-03