295
Configuring the CAN/LIN Definition Data Settings
Bit addressing
1.
Configure the least significant bit (LSB) of the first byte data in the data frame at address 0.
Allocate addresses to every bit toward the most significant bit (MSB), incrementing by one.
2. When the MSB is addressed, allocate the following addresses to the next byte data beginning
from the LSB toward the MSB, incrementing by one as with the first byte (byte 0).
3.
A data frame can have a size of eight bytes (64 bits) at most. The MSB of the eighth byte (byte 7)
is configured at address 63.
MSB
LSB
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
First
Byte 0
7
6
5
4
3
2
1
0
Byte 1
15
14
13
12
11
10
9
8
Byte 2
23
22
21
20
19
18
17
16
Byte 3
31
30
29
28
27
26
25
24
Byte 4
39
38
37
36
35
34
33
32
Byte 5
47
46
45
44
43
42
41
40
Byte 6
55
54
53
52
51
50
49
48
Last
Byte 7
63
62
61
60
59
58
57
56
To find out the data starting address (
[Start bit]
)
When you wish to capture 8-bit data of the second byte from the first (byte 1), configure the LSB at
address 8.
First of the data frame
Toward the last of the data frame
7 6 5 4 3 2 1 0 7 6 5 4 3 2 1
0
7 6 5 4 3 2 1 0 7 6 5 4 . . .
Byte 0
(1st byte)
Byte 1
(2nd byte)
Byte 2
(4th byte)
Byte 3
(2nd byte)
LSB position
Configured at address 8,
as shown by the above table.
CAN/LIN Measurement Function
HIOKI MR6000A966-03
Summary of Contents for Memory HiCorder MR6000
Page 2: ...HIOKI MR6000A966 03...
Page 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Page 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Page 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Page 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Page 391: ...HIOKI MR6000A966 03...
Page 392: ...HIOKI MR6000A966 03...