139
Triggering the Instrument Using Logic Signals (Logic Trigger)
3
Tap each of the signals under
[Trigger pattern]
to set a logic-trigger pattern.
X
Ignores a signal.
0
The logic-trigger condition for each logic signal is satisfied when the signal is at a low
level.
1
The logic-trigger condition for each logic signal is satisfied when the signal is at a high
level.
Setting example
The logic-trigger conditions differ depending on the combination of the
[Condition satisfied]
setting (logical OR or AND operation) and the
[Trigger pattern]
setting as follows:
Lx1 [1]
Lx2 [0]
Lx3 [X]
Lx4 [X]
Lx1 [1]
Lx2 [0]
Lx3 [X]
Lx4 [X]
OR
AND
Trigger pattern
Trigger pattern
Not satisfied
Satisfied
Not satisfied
Satisfied
Logic-trigger condition
Logic-trigger condition
Configuring the
Trigger Settings
HIOKI MR6000A966-03