168
Configuring the Numerical Calculation Settings
Setting
Description
Sample screen
Period
Frequency
Pulse width
Duty ratio
Target channel
(Analog, logic, real-time
waveform calculation)
Allows you to specify channels
as calculation targets.
Level
*
The instrument performs
calculations using values
based on a period of time
when a waveform crossed the
level specified here. No setting
is available for logic channels.
Slope
No slope can
be specified for
calculating the
duty ratio.
Performs calculations using
values based on a period of
time when a waveform crossed
the specified level while rising.
Performs calculations using
values based on a period of
time when a waveform crossed
the specified level while falling.
Filter(samples)
(
Off
,
10
to
10,000
)
Allows you to enter duration
used to determine that a
waveform crosses the specified
level.
Only after a waveform has
crossed the level and has
never crossed the level
again during the specified
filter duration, the instrument
determines that the waveform
has crossed the level.
This is useful to eliminate false
level-crossing events caused
by noise.
Stat.
Top
Detects a value obtained first
from the beginning within the
specified range.
Average
Calculates the average,
maximum, or minimum value
for each parameter within the
specified range.
Max
Min
Times
Calculates the value when the
number of times a waveform
crosses the level reaches the
specified number from the
beginning within the specified
range.
*: You can set a level within the range of −9.9999E+29 to −1.0000E29, at zero, within the range of +1.0000E29 to
+9.9999E+29. A value to five significant figures can be specified (a value to 10 significant figures can be specified
for a time of the level at time setting).
HIOKI MR6000A966-03
Summary of Contents for Memory HiCorder MR6000
Page 2: ...HIOKI MR6000A966 03...
Page 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Page 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Page 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Page 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Page 391: ...HIOKI MR6000A966 03...
Page 392: ...HIOKI MR6000A966 03...