157
Searching For Differences from a Fundamental Wave (Memory HiConcierge)
5
Tap the
[Mode]
box, and then from the list, choose a type to be searched for.
Histogram (comparison
with fundamental wave)
Searches for differences from the specified fundamental wave based on a
histogram.
Histogram (comparison
with previous waveform)
Searches for differences from the previous waveform based on a histogram.
Standard deviation
(comparison with
fundamental wave)
Searches for differences from the specified fundamental wave based on a
standard deviation.
Standard deviation
(Compared with previous
waveform)
Search for differences from the previous waveform based on a standard
deviation.
Memory HiConcierge searches a whole measured waveform for differences. You cannot specify a segment to
be searched.
6
Use the
[Period settings]
toggle switch to choose a period setting used for searches.
Auto
Automatically detects the period.
Any
Allows you to tap the
[Sample]
box to enter the number of samples, which specifies a
period.
With the
[Auto]
setting, the instrument may not be able to determine a period depending on measured
waveforms. If a fundamental wave have an unintended form, change this setting to
[Any]
and specify a period
in terms of the number of samples in the
[Sample]
box.
7
Tap
[Refresh fundamental wave]
to display a fundamental wave.
The instrument extracts a fundamental wave from the specified period and displays the wave on the screen.
8
Tap the
[Display order]
list and choose a display order from the list.
Sort by time
Displays retrieved positions in chronological order.
Sort by data
Displays retrieved positions in order of decreasing difference from a comparison target
(fundamental wave or previous waveform).
9
Tap
[Number to display]
box and enter the number of retrieved points to be displayed.
10
Tap
[Execute]
.
The instrument puts search marks ( ) at positions where a search condition is satisfied.
The retrieved position switching panel appear on the waveform screen.
To stop the searching
Press the
STOP
key.
Search Function
HIOKI MR6000A966-03
Summary of Contents for Memory HiCorder MR6000
Page 2: ...HIOKI MR6000A966 03...
Page 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Page 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Page 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Page 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Page 391: ...HIOKI MR6000A966 03...
Page 392: ...HIOKI MR6000A966 03...