External Input and Output
4
Tap
[External trigger]
.
The setting dialog box appears.
(2)
(3)
(1)
(1) Tap the
[External start trigger]
button to set it to
[On]
.
(2)
Tap the box to the right of the
[External start trigger]
box, and then from the list, choose which
direction to used for reception of the external trigger.
To accept a rising edge:
[ ]
*
To accept a falling edge:
[ ]
*
To accept both a rising edge and falling edge:
[ & ]
*
*: When the trigger satisfaction condition is set to
[AND]
, the instrument displays
[HIGH]
or
[LOW]
instead of them. You cannot set the trigger satisfaction condition to
[High or Low]
.
(3) Tap the
[Filter]
box, and then from the list, choose a filter setting.
This setting can prevent noise and chattering from triggering the instrument unintentionally.
Always set this setting to [Off] to synchronize multiple instruments using the external trigger.
Off
Enables the chattering prevention filter.
On
Disables the chattering prevention filter.
HIOKI MR6000A966-03