65
Configuring Module-Specific Settings
3.7
Configuring Module-Specific Settings
You can configure advanced settings for each module.
Configuring Model 8968 High Resolution Unit settings
> [Channel] > [8968]
1
Tap the area that includes
[A.A.F]
.
The
[A.A.F.]
setting dialog box appears.
2
Tap the
[A.A.F.]
box, and then choose
[On]
or
[Off]
from the list for the
anti-aliasing filter
setting.
Enabling the anti-aliasing filter prevents aliasing distortion from influencing FFT calculations.
The cutoff frequency automatically changes according to the sampling rate or frequency range (for the FFT
function) settings.
Off
Disables the anti-aliasing filter.
On
Enables the anti-aliasing filter.
(Disabled when the external sampling is used, or the sampling rate is set at 100 kS/s
or faster)
3
Tap
[Close]
.
The setting dialog box closes.
Advanced Functions
HIOKI MR6000A966-03