80
Configuring Module-Specific Settings
(When you have to choose a current sensor setting) Tap the
[Mode]
box, and then choose a
connected current sensor.
Available measurement ranges, which depend on chosen current sensors, are as follows:
CT7631/CT7731
200 A
CT7636/CT7736
200 A, 400 A, 1000 V
CT7642/CT7742
2000, 4000
CT7044/CT7045/
CT7046
2000, 4000, 10000
0.1mV/A
2000, 4000, 10000, 20000, 40000, 100000
1mV/A
200, 400, 1000, 2000, 4000, 10000
10mV/A
20, 40, 100, 200, 400, 1000
100mV/A
2, 4, 10, 20, 40, 100
1000mV/A
0.2, 0.4, 1, 2, 4, 10
IMPORTANT
When a current sensor included in Model CT6700 series connects via Model CT9920 Conversion
Cable, set a current sensor model name or its output rate.
2
(When you have changed the measurement mode) Tap
[Zero adjust]
.
The instrument performs zero-adjustment. Execute zero-adjustment without any input.
3
Tap the
[Range (f.s.)]
box, and then choose a measurement range from the list.
The instrument automatically configures the scaling setting for a measurement range according to a
recognized current sensor.
IMPORTANT
The figure of each measurement range name represents the maximum current that Model U8977
can measure using the range. However, the instrument cannot measure currents that exceed
the rated current of the connected current sensor. Check the specifications of the current sensor
used.
HIOKI MR6000A966-03
Summary of Contents for Memory HiCorder MR6000
Page 2: ...HIOKI MR6000A966 03...
Page 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Page 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Page 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Page 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Page 391: ...HIOKI MR6000A966 03...
Page 392: ...HIOKI MR6000A966 03...