Configuring Module-Specific Settings
Configuring Model 8970 Freq Unit settings
> [Channel] > [8970]
to
2
12
1
1
Tap the
[Mode]
box, and then choose a measurement mode from the list.
Freq
Measures frequency of a waveform (in hertz [Hz]).
Rotation speed
Measures the number of rotations of a measuring object (in rotations per minute
[r/min]).
Power frequency
Measures power frequency fluctuation (in hertz [Hz]).
Count
Accumulates the number of input pulses.
Duty ratio
Measures duty ratios of a waveform to be measured (in percent [%]).
Pulse width
Measures pulse widths (in second [s]).
A pulse that rises during the dead time (during a calculation) cannot be measured (with a frequency of 25 kHz
or higher).
Ignored
Calculation (40 μs)
Acquisition of the waveform
Acquisition of the waveform
2
Tap the area that includes
[Input voltage]
.
The setting dialog box appears.
3
Tap the
[Input voltage]
box, and then choose an option for the maximum level of an input
signal from the list.
±10 V
,
±20 V
,
±50 V
,
±100 V
,
±200 V
,
±400 V