3 • Operation and Maintenance
Doc. 065051-03 1/08
81
Example Simultaneous Injection Program: Two Valves Installed in
an ICS-1000/1500/2000
InjectValve_1.InjectPosition
;Switch the first inject
valve to Inject
InjectValve_2.InjectPosition
;Switch the second
inject valve to Inject
Inject
;Sweep the sample onto
the column
Wait
InjectState
;Wait for the injection to
finish
0.000
Pump_InjectValve.LoadPosition
;Switch the first inject
valve to Load
Pump_InjectValve_2.LoadPosition
;Switch the second
inject valve to Load
Load
;Load sample onto the
loop
Wait
CycleTimeState
;Wait for the end of the
specified Cycle Time
Pump_InjectValve.InjectPosition
;Switch the first inject
valve to Inject
Pump_InjectValve_2.InjectPosition
;Switch the second
inject valve to Inject
Inject
;Sweep the sample onto
the column
Wait
InjectState
;Wait for the injection to
finish
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...