5 • Service
Doc. 065051-03 1/08
159
Notes for calibrating the inject port volume when in sequential mode:
•
Preparing:
After disconnecting the diverter valve lines from the injection
valves, route the lines outside the autosampler or chromatography
compartment.
Make sure the sample syringe is free of bubbles (see
•
Step 1:
Select
Diverter Valve Position_1
to calibrate the inject port
volume for system 1.
Figure 5-28. Chromeleon Inject Port Volume Calibration Panel: Sequential Mode