B • Installation
Doc. 065051-03 1/08
195
Connecting to Two Valves Installed in the AS
1. Route the outlet lines from the pump and column in both
chromatography systems through the slot on the right side of the AS
(see
2. Connect the pump outlet line from system #1 to port P (2) on the
bottom injection valve (valve #1) and connect the column outlet line
from system #1 to port C (3) on valve #1 (see
3. Connect the pump outlet line from system #2 to port P (2) on the top
injection valve (valve #2) and connect the column outlet line from
system #2 to port C (3) on valve #2 (see
Figure B-28. Routing Pump and Column Lines to AS Injection Valves
Valve #2
Valve #1
Lines from column
and pump in
system #2
Lines from column
and pump in
system #1