[450] Non-recoverable motion error occurred in Y-axis. Clear any
physical obstructions and then press any key to reset system.
Possible Causes:
•
Vial obstructing motion
•
Misaligned sampling needle arm
Actions:
•
Remove the vial obstruction.
•
Align the sampling needle arm in the inject port (see
•
Turn off the power and check for freedom of movement of all
three axes of the sampling arm.
•
[451] Recoverable motion error occurred in Z-axis.
Possible Causes:
An error occurred during Z-axis movement of
the sampling arm. The DSP was able to recover from the error
and continue the motion.
Actions:
Occasional occurrences of this error can be ignored. If
the problem persists, contact Dionex.
•
[452] Non-recoverable motion error occurred in Z-axis. Clear any
physical obstructions and then press any key to reset system.
Possible Causes:
•
Vial obstructing motion
•
Bent sampling needle
•
Misaligned sampling needle arm
•
Incorrect X-Y position (not positioned above a vial or well)
Actions:
•
Remove the vial obstruction.
•
If the needle is bent, replace the sampling needle assembly (see
•
Align the sampling needle arm in the inject port (see
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...