AS Autosampler Operator’s Manual
270
Doc. 065051-03 1/08
C.9.7
Leak Sensor Calibration and Status Screen
Use this screen to display the status of the leak sensors and to calibrate
them.
Figure C-36. Leak Sensor Calibration and Status Screen
Screen Field
Description
MEASURED
VALUE
Reports the current measured voltages for the two leak sensors.
TRAY
is the sensor installed in the autosampler drip tray. The
OVEN
fields are for support of the thermal compartment option.
Note:
The thermal compartment option is not currently
available.
CURRENT
CONDITION
Reports the current state of the sensors:
DRY
if the sensor
reading is within the dry range,
WET
if the sensor reading is
within the wet range, or
ERR
if the sensor is out of range.
If the measured value drops 0.02 volts below the calibration
value, the current condition is wet. If the measured value goes
0.06 volts above the calibration value, the sensor needs to be
recalibrated.
To calibrate a sensor, press
Select
Δ
or
Select
∇
to select
CAL
and then press
Enter
. The current measured value becomes the
new calibration value. See
for detailed instructions.
CALIBRATION
VALUE
Reports the voltage values set for the calibration.
LOW LEAK
THRESHOLD
Reports the minimum voltage reading that is interpreted as a dry
sensor. A reading below this voltage indicates a wet sensor.
Help Prompt
LEAK SENSOR CALIBRATION AND STATUS
MEASURED VALUE:
TRAY
CURRENT CONDITION:
CALIBRATION VALUE:
LOW LEAK THRESHOLD:
2.49
2.29
DRY
2.48
ENABLE LEAK DETECTION YES
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...