3 • Operation and Maintenance
Doc. 065051-03 1/08
101
This message remains displayed until the sampling arm is in the home position
and it is safe to open the door. The screen then displays the message that it is safe
to open the door (see
The
Time
field counts down the amount of time remaining in which it is safe to
open the door. If you open the door, the sequence or schedule pauses. When you
close the door, the sequence or schedule resumes.
The timer continues counting down while the door is open. If the door is still open
when the timer reaches zero, the sequence is ended.
To cancel the
OPEN DOOR SAFETY
screen, press any front panel button. If the
door is open, a message cautions you that pressing a button again will end the
schedule.
Notes
•
When the AS is operating under computer control, the door can be opened
safely only when the sampling arm is in the home position and the
autosampler is idle. This can occur at various times in a program (for
example, after injection is complete, but before the next injection starts).
•
When the AS is operating under front panel control, the door can be opened
safely only when the sampling arm is in the home position and the status
displays
WAITING FOR CYCLE TIME TO EXPIRE
. The cycle time controls the
time between injections (see
). For most runs,
selecting a cycle time greater than 4 minutes provides an adequate wait period
between each injection for opening the door. If a method has a long sample
prep phase, however, a longer cycle time may be required.
Figure 3-38. Open Door Safety Screen: Safe to Open Door
NOTE: any key press will exit this mode
STATUS: WAITING
Time=5.47
OPEN DOOR SAFETY SCREEN
ATTENTION: It is now safe to open the
autosampler door. Please note time left
to safely open door.
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...