AS Autosampler Operator’s Manual
160
Doc. 065051-03 1/08
•
Step 2:
Clicking
Setup
empties the needle seal line and the diverter line.
•
Step 3:
Enter an approximate volume using the following formula:
0.8
μ
L/in (2
μ
L/in) for the diverter line plus the volume of the needle seal
line (which is written on a label on the tubing and inside the AS front
door). Then, click
Estimate
. The AS dispenses the volume entered, which
nearly fills the diverter and needle seal lines (minus a few microliters).
NOTE The value used for approximating the diverter line
volume (0.8
μ
L/cm, 2
μ
L/in) assumes a tubing ID of
0.330 mm (0.130 in). For volume per length values for
other tubing IDs, see
.
•
Step 4:
After clicking
+1 uL
,
wait until the sound of the syringe motor
stops before clicking
+1 uL
again.
Continue clicking
+1 uL
until a tiny drop of liquid appears at the end of
the diverter line. If you go too far and more than a single small drop of
liquid appears, click -
5 uL
. The AS will empty the line and then fill the
line with 5
μ
L less than was cleared. You can then begin clicking
+1 uL
again.
•
Step 5:
After clicking
Test
, verify that a drop of liquid appears at the end
of the diverter line. If no drop appears or if the drop is a different size than
previously observed, click -
5 uL
and repeat the procedure.
•
Step 6:
After clicking
Done
, the new calibration value is displayed on the
panel. Copy this value to the label inside the AS front door.
•
Calibrate System 2:
Select
Diverter Valve Position_2
and repeat the
calibration procedure for system 2.
•
Finishing:
Reconnect the diverter lines to the injection valves. Tighten
the fittings fingertight and then tighten an additional one-quarter turn
using the 5/16-in wrench.
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...