4 • Troubleshooting
Doc. 065051-03 1/08
109
•
Faulty cable
Actions:
•
Check the syringe motion. If the syringe action is faulty, the
syringe may need to be replaced (see
•
If the syringe does not move in response to a command, the cable
may need to be replaced. Contact Dionex.
•
[377] Time-out occurred waiting for syringe serial port txrdy.
Possible Causes:
•
Blocked syringe lines
•
Faulty syringe
•
Faulty cable
Actions:
•
Check the lines to the syringe for crimps or other blockages.
Replace the lines, if necessary (see
•
Check the syringe motion. If the syringe action is faulty, the
syringe may need to be replaced (see
•
If the syringe does not move in response to a command, the cable
may need to be replaced. Contact Dionex.
•
[378] Time-out occurred waiting for syringe to respond to
command.
Possible Causes:
Communication problem.
Actions:
Turn off the AS power, check the syringe cable
connections, and then turn on the power again. If the message
reappears, contact Dionex.
•
[379] Response from syringe contained data errors.
Possible Causes:
Communication problem.
Actions:
Turn off the AS power and then turn on the power again.
If the message reappears, contact Dionex.
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...