4. Route the green waste lines connected to port W (6) on each injection
valve through the side slots to the AS drip tray. Insert one line into the
small round opening in the tray (see
). Insert the other
line about 12 mm (1/2 in) into the large drain opening in the tray.
Figure B-29. Injection Valve Connections for Simultaneous Injection:
Valves Installed in the AS
Make sure the injection valve waste lines are not elevated above the
injection valves at any point between the valves and the AS drip tray.
Figure B-30. Injection Valve Waste Line Connections for Dual Valves
AS Needle
Seal Line
From Pump
in System #2
Sample In
To Waste
W
L
P
C
L
S
To Column
in System #2
Valve #2
From Pump
in System #1
To Waste
W
L
P
C
L
S
To Column
in System #1
Valve #1
Insert the waste
line from the first
injection valve
here
Insert the waste
line from the
second injection
valve here