C • Display Screens
Doc. 065051-03 1/08
249
C.6.3
Plumbing Configuration Screen
Use the
PLUMBING CONFIGURATION
screen to specify the plumbing
options installed in the
AS
.
Figure C-17. Plumbing Configuration Screen
Screen Field
Description
Values
LOOP SIZE
V1
The volume of the loop installed on the
first (or only) injection valve. The upper
limit depends on the volume of the sample
syringe installed:
500
μ
L (if the syringe volume is
≤
1 mL)
4,000
μ
L (if the syringe volume is 5 mL)
8,000
μ
L (if the syringe volume is 10 mL)
This field is not available in concentrate
mode.
2.0 to 99.9
μ
L
100 to x
μ
L
(default=25)
LOOP SIZE
V2
The volume of the loop installed on the
second injection valve. This field is
available in simultaneous and sequential
injection modes only.
•
In simultaneous mode (see
), the value entered for
LOOP SIZE V2
is for informational
purposes only.
•
), the value of
LOOP
SIZE V2
determines the injection type
). It is also used to
compute the injection volume.
See
LOOP SIZE V1
Help Prompt
PLUMBING CONFIGURATION
LOOP SIZE V1:
SAMPLE SYRINGE VOLUME:
PREP SYRINGE VOLUME:
SAMPLE SYRINGE CONFIG:
5
25
250
READY
uL
uL
mL
PREP SYRINGE CONFIG:
READY
uL
LOOP SIZE V2:
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...