AS Autosampler Operator’s Manual
286
Doc. 065051-03 1/08
Pipetting Liquid from One Vial to Another
The
PIPET
operation uses the sample syringe to transfer liquid from a
source vial to a destination vial. After pipetting, mix the destination vial
before sampling from it. Be sure to specify enough mix cycles to
completely mix the vial contents (see “Mixing a Vial” below).
NOTE Unlike the
DILUTE
operation, the
PIPET
operation does
not include an automatic mixing step. If the sample
preparation option is installed, use the
DILUTE
operation, not
PIPET,
).
Mixing a Vial
The
AS
mixes the vial by repeatedly drawing up and expelling a volume
of the contents. The prep syringe is used if the sample preparation option
is installed; otherwise, the sample syringe is used. When specifying the
volume, enter a value that is less than or equal to the volume of the
syringe. If the volume is greater than the syringe volume, an error
message occurs and the syringe volume is used.
20.0
uL
to
#S+1
#SMP
PIPET
Source Vial
Destination Vial
Volume to Pipet
Volume to Pipet
1 to 100
1 to 49
SMP, S+1, ..., S+9
(aluminum tray)
(plastic tray)
1 to 100
SMP, S+1, ..., S+9, WST
1 to 49
1.0 to 99.9 L
100 to 1000 L
μ
μ
Source Vial
Destination Vial
MIX
times
10
uL
250
#SMP
Vial to Mix
Vial to Mix
1 to 100 (aluminum tray)
1 to 49 (plastic tray)
SMP, S+1, ..., S+9
1 to 99
1 to 9999 L
μ
Number of
Mix Cycles
Cycles
Volume to
Aspirate and
Expel per Cycle
Volume
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...