AS Autosampler Operator’s Manual
162
Doc. 065051-03 1/08
5.11 Calibrating the Leak Sensor
If you are calibrating an existing (not a new) leak sensor, complete the following
steps before calibrating the leak sensor.
1. Remove the drip tray from the autosampler (see
) and thoroughly
clean and dry the tray and sensor.
2. Replace the sensor on the tray and reinstall the drip tray (see
The leak sensor can be calibrated from the front panel or from Chromeleon.
To calibrate the leak sensor from the front panel:
1. Press
Menu
,
8
, and
0
to go to the
LEAK SENSOR CALIBRATION AND STATUS
screen.
Figure 5-29. Leak Sensor Calibration and Status Screen
2. Press
Select
Δ
or
Select
∇
to select
CAL
and then press
Enter
.
The current measured value becomes the new calibration value.
To calibrate the leak sensor from the Chromeleon Wellness panel:
1. Open the Chromeleon Browser and expand the
Dionex Templates
folder.
2. Expand the
Panels
folder and the
Wellness
folder.
3. On the right pane of the Browser window, double-click
Dionex_autosampler_wellness.pan
. The Wellness panel appears.
Help Prompt
LEAK SENSOR CALIBRATION AND STATUS
MEASURED VALUE:
TRAY
CURRENT CONDITION:
CALIBRATION VALUE:
LOW LEAK THRESHOLD:
2.49
2.29
DRY
2.48
ENABLE LEAK DETECTION YES
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...