8. Connect the needle seal line to port S on the second injection valve in
the DC.
9. Locate the green waste line connected to port W (6) on the injection
valve (see
10. Route the waste line from the injection valve through the side slot on
the DC and then through the side slot on the AS.
11. To ensure the green waste line stays in place in the opening in the AS
drip tray, sharply bend the line about 12 mm (1/2 in) from the end,
being careful not to restrict the inner diameter.
12. Push the bent end of the line into the small round opening on the right
side of the
AS
13. Calibrate the inject port volume. Refer to
Make sure the injection valve waste line is not elevated above the
injection valve at any point between the valve and the AS drip tray.