4 • Troubleshooting
Doc. 065051-03 1/08
119
•
[445] Invalid leak sensor reading. May be open- or short-circuit, or
bad sensor. See Leak Sensor Calibration screen for more
information.
Actions:
The leak sensor card, the cables, or the sensor must be
replaced. Contact Dionex.
•
[446] Current leak sensor voltage is outside of legal calibration
range.
Attempting to calibrate when the leak sensor voltage reading is
outside the normal dry range.
Possible Causes:
The sensor is not dry enough for calibration.
Actions:
Clean and dry the area thoroughly and recalibrate (see
•
[447] Leak sensor needs recalibration. Go to Leak Sensor
Calibration screen to perform recalibration operation.
Actions:
See
•
[448] Motion command to DSP was aborted. Needle may not have
homed properly.
Possible Causes:
Loose coupling in the Z-axis. The needle does
not stay up.
Actions:
Try the motion again. If the problem persists, contact
Dionex.
•
[449] Recoverable motion error occurred in Y-axis.
Possible Causes:
An error occurred during Y-axis movement of
the sampling arm. The DSP was able to recover from the error
and continue the motion.
Actions:
Occasional occurrences of this error can be ignored. If
the problem persists, contact Dionex.
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...