AS Autosampler Operator’s Manual
290
Doc. 065051-03 1/08
Reagent Flush after Concentrator Loading or Reagent Priming
REAGENT FLUSH
is used to flush reagent onto the concentrator column
for details about concentrator
injections.
The following steps occur during reagent flush:
•
(Optional) The inject valve is switched to a specified position (load or
inject). For most applications, the valve position is not changed.
•
If the reagent source is a vial, the needle arm moves to the vial and
picks up the volume of reagent (see the notes below for volume
ranges and defaults).
•
The needle arm moves to the inject port.
•
The syringe delivers the specified volume of reagent from the
reservoir to the inject port. From there it moves onto the concentrator.
The sample syringe and the flush reservoir are used by default for
reagent flushing. If the sample prep option is installed, the prep
syringe is used and a reagent reservoir can be specified.
Notes on reagent prime and flush volume ranges:
•
If the source is a reservoir, the allowed range is 100 to 32,600
μ
L. The
default is 2,000
μ
L if the syringe volume is less than 5,000
μ
L, and
10,000
μ
L if the syringe volume is 5,000 or 10,000
μ
L.
1000 uL
Vlv
FLH
N
RGNT_FLH
Reagent Source
Inject Valve Position
Volume to Flush
Volume
Vial or flush
reservoir. Also,
RES A, B, C, or D,
if sample prep is
installed
L (load)
I (inject)
N (no change)
100.0 to
50,000 L
(see notes)
μ
Source
Inject Valve
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...