AS Autosampler Operator’s Manual
28
Doc. 065051-03 1/08
Set these functions from the
DETAIL STATUS
screen (see
Local/Schedule Control Mode
The AS runs all injections according to a programmed schedule. The
schedule defines a series of vials to be sampled and the operating
parameters to run on each vial. A schedule consists of the following
parameters:
•
The vials to sample
•
The number of injections to draw from each vial
•
The volume of each injection
•
The method to run on each injection
•
The action to take when a scheduled vial is missing
You program, save, and edit schedules from the
SCHEDULE
screen (see
). The AS can store up to nine schedules in memory. The
stored schedules are retained in memory when the AS is turned off.
describes how to create schedules.
Figure 2-11. Detail Status Screen Example
Figure 2-12. Schedule Screen
Help Prompt
IDLE
VIAL#:
TIME:
INJ#:
VOL:
METHOD:
TRAY:
TRAY:
LOOP:
20 °C
25
100
2 mL
uL
uL
min
TTL1
TTL2
RLY1
RLY2
0
1
0
0
LOCAL
DETAIL
DIRECT CONTROL
INJ VLV L
NORMAL FULL
Help Prompt
SAVE TO
RUN
SCHEDULE
25.0
1
1
1
1
MISSING VIAL ACTION:
INJ/
INJ
VIAL
METHOD
VOL(uL)
VIAL#
LINE
END
START
1
EDIT
STOP
-
-
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...