[380] Response from syringe was too long. Receive buffer
overflowed.
Possible Causes:
Communication problem.
Actions:
Turn off the AS power and then turn on the power again.
If the message reappears, contact Dionex.
•
[381] An attempt was made to move the syringe to an illegal
position. Most likely, the specified volume was larger than the
remaining syringe capacity.
Valid positions for the syringes are programmed into the AS
instrument control Moduleware. There was an internal instruction to
move a syringe to a position that was not programmed.
Possible Causes:
Problem with the Moduleware.
Actions:
Turn off the AS power and then turn on the power again.
If the message reappears, contact Dionex.
•
[382] An attempt was made to move a syringe valve to an illegal
position.
Valid positions for the syringe valves are programmed into the AS
instrument control Moduleware. There was an internal instruction to
move a syringe valve to a position that was not programmed.
Possible Causes:
Problem with the Moduleware.
Actions:
Contact Dionex.
•
[383] An attempt was made to set the syringe speed to an illegal
value.
Valid speeds for the syringes are programmed into the AS instrument
control Moduleware. There was an internal instruction to move a
syringe at a speed that was not programmed.
Possible Causes:
Problem with the Moduleware.
Actions:
Contact Dionex.
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...