3.6.2
Determining the Injection Type
To determine the injection type performed by the AS, first select the
sample mode on the
SYSTEM PARAMETERS
Press
Menu
,
5
, and
5
to go to the
SYSTEM PARAMETERS
screen.
The sample mode corresponds to the AS configuration. The following
sample modes are available:
•
Normal
—Sample is delivered to a single system; a sample loop is
installed on the injection valve.
•
Concentrate
—Sample is delivered to a single system; a concentrator
column is installed on the injection valve.
•
Simultaneous
—Sample is delivered to two systems simultaneously;
a sample loop is installed on both injection valves.
•
Sequential
—Sample is delivered to two systems alternately; a
sample loop is installed on both injection valves.
•
Sequential concentrate
—Sample is delivered to two systems
alternately; a concentrator column is installed on both injection
valves.
NOTE The
SAMPLE MODE
must be selected on the
SYSTEM PARAMETERS
screen. It cannot be selected
from Chromeleon or Chromeleon Xpress.
If normal or sequential mode is selected, you must also enter additional
parameters to specify the type of sample loop injection performed (see
Figure 3-14. System Parameters Screen
Help Prompt
SYSTEM PARAMETERS
WAIT:
ENABLED
METHOD SYRINGE SPEED:
CUT SEGMENT VOLUME:
INJECT PORT VOLUME:
uL
uL
10
S1 43
3
SAMPLE MODE:
NORMAL
FLUSH SYRINGE SPEED:
4
uL
S2
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...