4 • Troubleshooting
Doc. 065051-03 1/08
121
•
Turn off the power and check for freedom of movement of all
three axes of the sampling arm.
•
Contact Dionex.
•
[453] Recoverable motion error occurred in inject valve.
Possible Causes:
•
Valve in wrong position
Actions:
Turn off the AS power and then turn on the power again. If
the message reappears, contact Dionex.
•
[454] Non-recoverable motion error occurred in inject valve. Clear
any physical obstructions and then press any key to reset
system.
Possible Causes:
Valve in wrong position
Actions:
Turn off the AS power and then turn on the power again. If
the message reappears, contact Dionex.
•
[455] Recoverable motion error occurred in auxiliary valve.
Possible Causes:
Valve in wrong position
Actions:
Turn off the AS power and then turn on the power again. If
the message reappears, contact Dionex.
•
[457]-[474] Syringe
reports fatal error:
Syringe reports error:
Possible Causes:
Syringe errors can result from various
mechanical problems or from a communication failure between
the syringe and the AS electronics.
Actions:
•
Check the syringe cable connections.
•
For fatal errors, press a key to reset the system.
•
If the error persists, contact Dionex.
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...