AS Autosampler Operator’s Manual
312
Doc. 065051-03 1/08
NOTE This procedure has numerous steps. To help maintain
your place in the procedure, check off each step after
completing it.
1. Disconnect the needle seal line fitting from port S (5) of the injection valve.
For sequential mode, disconnect the diverter valve line fitting from port S (5)
on each injection valve. For simultaneous mode, disconnect the needle seal
line fitting from the Y-connector.
2. Press
Menu
,
8
, and
5
to go to the
LIQUID CONTROL
screen.
3. Select
VIAL#
:
INJ
and press
Enter
.
The needle arm moves to the inject port.
NOTE Throughout the procedure, retain the current or default
settings for parameters, unless instructed otherwise.
4. Select the following options:
from
FLUSH
SYRINGE SPEED
:
5
SYRINGE
:
SAMPLE
ACTION
:
FILL
5. Press
Enter
.
NOTE To initiate a selected action, the cursor must be in the
ACTION
field when you press
Enter
.
6. Select
ACTION:
EMPTY
and press
Enter
.
Figure D-19. Liquid Control Screen
Help Prompt
LIQUID CONTROL
from
VIAL#:
ASPIRATE:
SYRINGE:
ACTION:
DISPENSE:
1
100
100
SAMPLE
ASPIRATE
uL
uL
NEEDLE
LOAD
46
NEEDLE HEIGHT:
SYRINGE SPEED: 4
INJ VALVE:
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...