To manually flush the inject port from the front panel:
1. Set the
AS
to
LOCAL
mode and
DIRECT CONTROL
.
2. Go to the
FLUSH/PRIME
screen
(press
Menu
and
6
)
and enter the
desired
FLUSH VOLUME
, or keep the default value.
3. Move the cursor to the
ACTION
field and select
FLUSH
(see
).
Figure 3-39. Initiating a Manual Flush Cycle
4. Press
Enter
to start the flush.
3.11 System Shutdown
No special system shutdown procedure is required. If the AS is under Chromeleon
or Chromeleon Xpress control, you can load a shutdown program at the end of the
schedule to automatically shut down the system. The shutdown program turns off
the pump flow, the suppressor current, the sample temperature control option (if
installed), and other system devices.
If you are running Chromeleon, you can use the SmartShutdown Wizard to assist
you in creating and running a shutdown program. Refer to the Chromeleon Help
for details.
Help Prompt
FLUSH/PRIME
VOLUME:
RESERVOIR:
ACTION:
uL
400
FLUSH
FLUSH
SYRINGE:
PRIME
FLUSH
VOLUME:
uL
400
RESERVOIR: RES A
SYRINGE:
RES A
PREP
ACTION: READY
SAMPLE
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...