D • Front Panel Control
Doc. 065051-03 1/08
305
Selecting the Method or Flush Syringe Speed
1. Go to the
SYSTEM PARAMETERS
screen (press
Menu
,
5
, and
5
).
2. Position the cursor in the
METHOD
SYRINGE SPEED
field.
3. Press
Select
Δ
or
Select
∇
to select the desired speed or enter the speed by
pressing a numeric button. 1 is the slowest speed and 5 is the fastest. The
default speed is 3, which corresponds to the speed required for water. Select a
slower setting for more viscous samples.
In addition, be sure to select a speed that will keep the backpressure below
690 kPa (100 psi), taking into account the pressure from the tubing and the
concentrator column (if installed).
See
for details.
4. Position the cursor in the
FLUSH SYRINGE SPEED
field. Press
Select
Δ
or
Select
∇
to select the desired speed or enter the speed by pressing a numeric
button (1 through 5). The default speed is 4.
Figure D-14. System Parameters Screen
Help Prompt
SYSTEM PARAMETERS
WAIT:
ENABLED
METHOD SYRINGE SPEED:
CUT SEGMENT VOLUME:
INJECT PORT VOLUME:
uL
uL
10
S1 43
3
SAMPLE MODE:
NORMAL
FLUSH SYRINGE SPEED:
4
uL
S2
Summary of Contents for As
Page 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Page 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Page 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Page 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Page 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Page 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...